Model Number SV-2101
Device Problems Obstruction of Flow (2423); Device-Device Incompatibility (2919)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/11/2023
Event Type  malfunction  
Manufacturer Narrative
Impact code f05 is being used to capture the reportable issue of aborted/cancelled procedure.
 
Event Description
It was reported to boston scientific corporation that a spaceoar vue device was going to be used in a spaceoar vue implant procedure, performed with general anesthesia, on (b)(6), 2023.During insertion, the visualization on the ultrasound was disrupted due to air from the syringes.The physician attempted to re-find the tip of the injection needle.However, by the time the needle was situated in the correct position, the kit clogged.As a result, the physician decided to abort the procedure for the patient's safety.There were no patient complications as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPACEOAR VUE SYSTEM
Type of Device
ABSORBABLE PERIRECTAL SPACER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
500 commander shea boulevard
quincy MA 02171
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key16298482
MDR Text Key308963423
Report Number3005099803-2023-00439
Device Sequence Number1
Product Code OVB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K182971
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSV-2101
Device Catalogue NumberSV-2101
Device Lot Number0030712569
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/11/2023
Initial Date FDA Received02/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/10/2022
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient SexMale
-
-