• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD10; SYSTEM, X-RAY, ANGIOGRAPHIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD10; SYSTEM, X-RAY, ANGIOGRAPHIC Back to Search Results
Model Number ALLURA XPER FD10
Device Problem Unintended System Motion (1430)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/23/2023
Event Type  malfunction  
Event Description
It has been reported to philips that the c-arc would not stop immediately when moving.The device was inside clinical use at the time of the event.No patient harm was reported.A philips engineer visited site and replaced the geometry module.The device was returned to expected functionality.
 
Manufacturer Narrative
Philips has investigated this complaint.The philips field service engineer (fse) inspected the system onsite and confirmed that the c-arc was running in certain angles and does not stop.Upon functional testing, the fse determined that the geometry control panel was defective and joystick was stiff on one side.As a part of repair activity, the fse replaced the geometry control panel.After replacement, the system was returned to use in good working order.The codes were updated based on the investigation outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD10
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
dusty leppert
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key16299487
MDR Text Key308846459
Report Number3003768277-2023-00733
Device Sequence Number1
Product Code IZI
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K041949
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Literature,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA XPER FD10
Device Catalogue Number722003
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/23/2023
Initial Date FDA Received02/03/2023
Supplement Dates Manufacturer Received11/09/2023
Supplement Dates FDA Received11/29/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Reuse
Patient Sequence Number1
-
-