• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DREAMSTATION CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DREAMSTATION CPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DSX200H11
Device Problems Contamination (1120); Degraded (1153)
Patient Problems Excessive Tear Production (2235); Sore Throat (2396); Respiratory Tract Infection (2420); Unspecified Respiratory Problem (4464)
Event Date 08/03/2021
Event Type  malfunction  
Event Description
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap/ bipap, and mechanical ventilator devices.The manufacturer received information alleging sinus irritation, congestion, watery eyes and throat irritation.There was no report of patient harm or injury.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Manufacturer Narrative
The manufacturer was previously contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap/ bipap, and mechanical ventilator devices.The manufacturer received information alleging sinus irritation, congestion, watery eyes and throat irritation.There was no report of patient harm or injury.The device was returned to the manufacturer's product investigation laboratory for investigation.An internal visual inspection of the device was completed by the manufacturer.The manufacturer found evidence of dust contamination on the blower and on the blower impeller, and water ingress on the bottom of the blower box, on the blower, and near the heater plate.The manufacturer found evidence of sound abatement foam degradation/breakdown.The humidifier was returned to the manufacturer's product investigation laboratory for evaluation.An internal visual inspection was completed by the manufacturer.The manufacturer found no evidence of contamination.The device's event logs were downloaded and reviewed.The manufacturer found 1 instances error logged.The manufacturer concludes there was evidence multiple contamination from the device.The manufacturer confirmed there was evidence of sound abatement foam degradation/breakdown.Humidifier: dsxhcp (h188610281ac7).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION CPAP
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key16299496
MDR Text Key308839197
Report Number2518422-2023-03798
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDSX200H11
Device Catalogue NumberDSX200H11
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/30/2023
Initial Date Manufacturer Received 08/03/2021
Initial Date FDA Received02/03/2023
Supplement Dates Manufacturer Received05/23/2023
Supplement Dates FDA Received06/23/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/18/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberRES 88058
Patient Sequence Number1
-
-