Model Number 0998-00-3023-53
Device Problem Circuit Failure (1089)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/24/2023
Event Type  malfunction  
Event Description
It was reported that the cs300 intra-aortic balloon pump has screen issues.
 
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported prior to use the cs300 intra-aortic balloon pump has screen was very dim.There was no patient involvement.
 
Manufacturer Narrative
It was reported that the cs300 intra-aortic balloon pump (iabp) while turning on the pump and the screen was very dim, over about 30 seconds the screen was fully lit up.The customers were concerned that the console may not function properly during therapy so they opted to switch it out and would like service to check it to be safe.A getinge field service engineer fse was dispatched to the site and evaluated the unit.He could not duplicate the issue while servicing.Pm complete per factory specifications.Calibrations, functional testing, and safety testing all passed per factory specifications.Device returned to customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key16302808
MDR Text Key308841467
Report Number2249723-2023-00959
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 02/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/24/2023
Initial Date FDA Received02/06/2023
Supplement Dates Manufacturer Received02/21/2024
Supplement Dates FDA Received02/22/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/08/2007
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-