Model Number CLK-4
Device Problem Optical Problem (3001)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/04/2023
Event Type  malfunction  
Event Description
The customer reported the halogen light source fails to light up.The bulb was replaced; however, the issue persisted.The event occurred during preparation for use and there was no patient harm or user injury reported due to the event.
 
Manufacturer Narrative
The device was returned for investigation.Upon evaluation of the device, the reported issue was confirmed by faulty power switch.In addition, front panel scratch, front socket air jointer worn-out, and power inlet crack were observed.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time, however, if additional information becomes available, this report will be supplemented accordingly.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.As a result of checking complaints in the past one year from the occurrence date of this complaint, the subject device had no repair history.Since the subject device was manufactured over 15 years ago, its record could not be checked.Based on the results of the investigation, a definitive root cause could not be established.The suggested phenomenon was presumed to have been due to a failure of the power switch.As to the cause of the defect, it was not possible to further identify it because no further information could be obtained.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HALOGEN LIGHT SOURCE
Type of Device
HALOGEN LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16308566
MDR Text Key309162661
Report Number3002808148-2023-01116
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170024528
UDI-Public04953170024528
Combination Product (y/n)N
Reporter Country CodeEZ
PMA/PMN Number
K993041
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLK-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/17/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/10/2023
Initial Date FDA Received02/06/2023
Supplement Dates Manufacturer Received03/30/2023
Supplement Dates FDA Received04/24/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-