This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, g3, g6, h1, h2, h3, h6, h10 h6: proposed component code: mechanical (g04)- drill reported event was unable to be confirmed due to limited information received from the customer.Visual evaluation of the returned device shows drill pin had fractured off and a piece remains lodged in one of the holes.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|