Device Problems Use of Device Problem (1670); Improper or Incorrect Procedure or Method (2017)
Patient Problem Insufficient Information (4580)
Event Date 01/11/2023
Event Type  malfunction  
Event Description
It was reported that there is a clinician misuse issue, stating the procedure for the ciaglia technique is "almost never followed".The clinician began to make a horizontal or vertical incision instead of starting to identify and mark the anatomical landmarks.It was reported that the clinician reversed steps 3 and 4 in the ifu.There has been no report of patient injury or no observable clinical symptoms or a change in symptoms identified in the patient.
 
Manufacturer Narrative
Catalog number, lot number, expiration date, and udi section and device manufacture date are unknown, no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when required.
 
Manufacturer Narrative
Other text: h6: event methods, evaluation, and conclusion codes: updated.No product was returned for analysis.No product lot number was provided, therefore, no review of manufacturing device history records could be conducted.Based on the available information, the reported issue could not be confirmed and no root cause could be found.No actions have been taken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX TRACHEOSTOMY CARE KIT
Type of Device
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
boundary road
hythe, kent CT21 6JL
UK   CT21 6JL
Manufacturer Contact
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key16311151
MDR Text Key308932421
Report Number3012307300-2023-00936
Device Sequence Number1
Product Code BTO
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K862167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/16/2023
Initial Date FDA Received02/07/2023
Supplement Dates Manufacturer Received02/21/2023
Supplement Dates FDA Received03/08/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-