Model Number MT24078 |
Device Problem
Unable to Obtain Readings (1516)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/16/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that error icon displayed occurred.No product or data was provided for evaluation.The allegation and a probable cause could not be determined.No injury or medical intervention was reported.
|
|
Event Description
|
The product was evaluated.An external visual inspection was performed and passed.Functional tests were not performed due to the receiver not turning on.A receiver charge test was performed and failed as the receiver did not turn on, but the receiver charged when using a known good battery.A receiver boot test was performed and passed.An internal visual inspection was performed and passed.The receiver log was downloaded and reviewed finding error related to the complaint.The allegation was confirmed.The probable cause was determined to be a deflective battery.No injury or medical intervention was reported.
|
|
Manufacturer Narrative
|
(b)(4).B5: describe event or problem - additional.D9: device availability - additional.H2: additional information/device evaluation.H3: device evaluated by manufacturer - additional.H6: adverse event problem - additional.
|
|