Model Number 8506Y
Device Problem Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
The customer reported that the staff had a near miss (needle stick) when trying to push the vacutainer in.The container that was not overfilled.
 
Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Manufacturer Narrative
Samples were not received for the investigation.A device history record review could not be performed because a lot number was not received with the complaint.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.Because a sample was not returned, we were unable to perform a follow up investigation to include functional and visual evaluations to confirm the reported issue and determine the root cause, therefore a corrective action is not applicable at this time.If a sample is received at a later date, this complaint will be reopened for further investigation.This complaint will be used for tracking and trending purposes.
 
Event Description
The customer reported that the staff had a near miss when trying to push the vacutainer in.The container that was not overfilled.Additional information received on 08feb2023 the customer stated that when you put blood culture vacutainer adaptors into these sharps containers, they actually don¿t fit, and block it from opening/closing for future sharps.Essentially it means you can¿t put these in or it will mean even if the container isn¿t full it will need to be replaced early.
 
Manufacturer Narrative
Added information to section b5 and e1 based on additional details provided by the initial reporter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5QT.CNT.W/TRNS YELLOW CONT. 20
Type of Device
CONTAINER, SHARPS
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key16315285
MDR Text Key309255171
Report Number1282497-2023-10734
Device Sequence Number1
Product Code MMK
UDI-Device Identifier10884521023260
UDI-Public10884521023260
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 02/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8506Y
Device Catalogue Number8506Y
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/23/2023
Initial Date FDA Received02/07/2023
Supplement Dates Manufacturer Received01/23/2023
01/23/2023
Supplement Dates FDA Received02/09/2023
02/24/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage A
Patient Sequence Number1
-
-