MAUDE Adverse Event Report: C.R. BARD INC. (COVINGTON) -1018233 BARDIA® FOLEY CATHETER SILICONE COATED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD INC. (COVINGTON) -1018233 BARDIA® FOLEY CATHETER SILICONE COATED Back to Search Results
Catalog Number 123520A
Device Problem Dent in Material (2526)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/20/2023
Event Type  malfunction  
Event Description
It was reported that foley catheter balloon was not symmetrical.The user reinserted a new catheter.
 
Manufacturer Narrative
The reported event was inconclusive ¿ poor sample condition.The reported failure was able to be reproduced.It is unknown whether the device had met relevant specifications.The product was used for urological care.Based on the photos, it was observed that the catheter balloon was asymmetrically inflated.However, it is unknown if the balloon was under inflated which could also cause the asymmetrical balloon issue.Besides, no functional testing could be performed as only photo sample is provided.Therefore, the reported event is inconclusive due to poor sample condition.The potential root cause for this failure mode could be user related (example: under-inflation during use)/ assembly of the balloon to the shaft/ uneven thickness of balloon.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "bard® urological catheter warning: do not use ointments or lubricants having a petrolatum base.They will damage latex.Units this is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Visually inspect the product for any imperfections or surface deterioration prior to use.Use luer tip syringe to inflate with stated ml of sterile water.Or for pre-filled products, remove clip and squeeze reservoir to inflate with stated ml of sterile water.For urological use only.To deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Female use only syringe of sterile water for balloon inflation refer to direct unit label for product content and gender specific use where applicable.Contains or presence of natural rubber latex.Caution: this product contains natural rubber latex which may cause allergic reactions.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Refer to direct unit label for sterilization method utilized sterilized by radiation sterilized by ethylene oxide warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable laws and regulations.".
 
Event Description
It was reported that foley catheter balloon was not symmetrical.The user reinserted a new catheter.
 
Manufacturer Narrative
The reported event was inconclusive ¿ poor sample condition.The reported failure was able to be reproduced.It is unknown whether the device had met relevant specifications.The product was used for urological care.Based on the photos, it was observed that the catheter balloon was asymmetrically inflated.However, it is unknown if the balloon was under inflated which could also cause the asymmetrical balloon issue.Besides, no functional testing could be performed as only photo sample is provided.Therefore, the reported event is inconclusive due to poor sample condition.The potential root cause for this failure mode could be user related (example: under-inflation during use)/ assembly of the balloon to the shaft/ uneven thickness of balloon.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "warning: do not use ointments or lubricants having a petrolatum base.They will damage latex.Visually inspect the product for any imperfections or surface deterioration prior to use.¿ use luer tip syringe to inflate with stated ml of sterile water.Or ¿ for pre-filled products, remove clip and squeeze reservoir to inflate with stated ml of sterile water.For urological use only.To deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.This is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDIA® FOLEY CATHETER SILICONE COATED
Type of Device
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key16323786
MDR Text Key309141487
Report Number1018233-2023-00704
Device Sequence Number1
Product Code EZC
Combination Product (y/n)N
Reporter Country CodeSN
PMA/PMN Number
K910846
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number123520A
Device Lot NumberMYFY3476
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/20/2023
Initial Date FDA Received02/08/2023
Supplement Dates Manufacturer Received09/21/2023
Supplement Dates FDA Received09/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-