Model Number 0570-0404
Device Problem No Visual Prompts/Feedback (4021)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/18/2023
Event Type  malfunction  
Manufacturer Narrative
The customer stated the "issue had been resolved," and declined the option to have their glidescope core 15-inch monitor returned to verathon for evaluation.The customer did not provide the solution that resolved the issue, just that the initial reported issue had stopped.Since the reported device was not returned to verathon for evaluation, the cause could not be determined.Review of complaint history for the reported glidescope core 15-inch monitor serial number "cn203168" did not identify any previous complaints reported to verathon.Trending analysis for the glidescope core 15-inch monitor does not identify any trends exceeding acceptable limits.Review of the system risk assessment confirmed that the risk associated with the hazardous scenario has been adequately captured and characterized by verathon.Corrective action is currently not required at this time.Verathon will continue to monitor for any ongoing trends.
 
Event Description
The customer reported that during a patient procedure, using a glidescope core 15-inch monitor, the image would freeze requiring the end user to reboot the video monitor.No reported use of a backup device or harm to the patient was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE CORE 15-INCH MONITOR
Type of Device
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA  V5C 5A9
Manufacturer (Section G)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA   V5C 5A9
Manufacturer Contact
corey kasbohm
20001 n creek pkwy
bothell, WA 98011-8218
4256295760
MDR Report Key16325670
MDR Text Key309155042
Report Number9615393-2023-00023
Device Sequence Number1
Product Code CCW
UDI-Device Identifier00879123006639
UDI-Public010087912300663911201124
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 01/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0570-0404
Device Catalogue Number0270-1033
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/18/2023
Initial Date FDA Received02/08/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/24/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-