• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL CADD EXTENSION SETS; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL CADD EXTENSION SETS; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-7047-24
Device Problems Fluid/Blood Leak (1250); Leak/Splash (1354); Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/11/2023
Event Type  malfunction  
Event Description
It was reported the disposable leaked chemo on the pump.They sent home a self-contained bag so the patient was not affected.No patient injury reported.
 
Manufacturer Narrative
A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
Other, other text: device evaluation: received two samples for investigation; the returned samples were received inside a plastic bag which it is not its original packaging.No obstructions, damaged, broken, or other defects were detected in none of the joins of the product.The sample returned was tested using the hydrostatic vessel.No leaks were found fleeing from the sample, thus the failure mode reported is not confirmed.Other analysis was not required.No mitigation apply since complaint was not confirmed.Root cause cannot be determined since the complaint was not confirmed.No corrective actions are required since the complaint was not confirmed.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD EXTENSION SETS
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key16329372
MDR Text Key309157456
Report Number3012307300-2023-01053
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10610586020315
UDI-Public10610586020315
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K942046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number21-7047-24
Device Catalogue Number21-7047-24
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/12/2023
Initial Date FDA Received02/08/2023
Supplement Dates Manufacturer Received03/16/2023
Supplement Dates FDA Received03/22/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-