• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION VERCISE CARTESIA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION VERCISE CARTESIA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS Back to Search Results
Model Number DB-2202-45
Device Problem High impedance (1291)
Patient Problems Hemorrhage/Bleeding (1888); Cognitive Changes (2551); Movement Disorder (4412); Swelling/ Edema (4577)
Event Date 01/12/2023
Event Type  Injury  
Event Description
It was reported that the deep brain stimulation (dbs) patient was hospitalized for potential swelling around the leads that resulted in gait issues.A computed tomography (ct) was performed, and no bleeds were detected from the leads.The patient remained in the hospital for observation.The patient underwent a magnetic resonance imaging scan (mri) which showed a little swelling and bleeding around the left electrode.The patient stated he was getting better each day but still had a slight gait issue and cognitively did not feel back to normal.An impedance check indicated that the left electrode had higher than normal impedances which the physician felt was due to the swelling and bleeding that occurred after implant.The physician believed the impedances would drop once the swelling and fluid eased up.The patient had the dbs system activated and was released from the hospital.No further action is planned.
 
Manufacturer Narrative
Additional suspect medical device component involved in the event: product family: dbs-linear leads, upn: m365db2202450, model: db-2202-45, serial: (b)(4), batch: 7097788.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERCISE CARTESIA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
erik sherburne
25155 rye canyon loop
valencia, CA 91355
7632920920
MDR Report Key16329776
MDR Text Key309123798
Report Number3006630150-2023-00495
Device Sequence Number1
Product Code NHL
UDI-Device Identifier08714729905288
UDI-Public08714729905288
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberDB-2202-45
Device Catalogue NumberDB-2202-45
Device Lot Number7097092
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/17/2023
Initial Date FDA Received02/08/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/06/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age70 YR
Patient SexMale
-
-