BOSTON SCIENTIFIC NEUROMODULATION VERCISE CARTESIA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS
|
Back to Search Results |
|
Model Number DB-2202-45 |
Device Problem
High impedance (1291)
|
Patient Problems
Hemorrhage/Bleeding (1888); Cognitive Changes (2551); Movement Disorder (4412); Swelling/ Edema (4577)
|
Event Date 01/12/2023 |
Event Type
Injury
|
Event Description
|
It was reported that the deep brain stimulation (dbs) patient was hospitalized for potential swelling around the leads that resulted in gait issues.A computed tomography (ct) was performed, and no bleeds were detected from the leads.The patient remained in the hospital for observation.The patient underwent a magnetic resonance imaging scan (mri) which showed a little swelling and bleeding around the left electrode.The patient stated he was getting better each day but still had a slight gait issue and cognitively did not feel back to normal.An impedance check indicated that the left electrode had higher than normal impedances which the physician felt was due to the swelling and bleeding that occurred after implant.The physician believed the impedances would drop once the swelling and fluid eased up.The patient had the dbs system activated and was released from the hospital.No further action is planned.
|
|
Manufacturer Narrative
|
Additional suspect medical device component involved in the event: product family: dbs-linear leads, upn: m365db2202450, model: db-2202-45, serial: (b)(4), batch: 7097788.
|
|