Model Number N/A |
Device Problem
Material Twisted/Bent (2981)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/18/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the tip of the anchor was slightly bent.Doctor attempt to straighten it with a needle driver but implant didn't work properly.Attempts have been made and there is no further information at this time.
|
|
Manufacturer Narrative
|
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the device has been discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
(b)(4).Reported event was confirmed as visual examination of the provided pictures identified the nitinol pusher had bent.Device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|