Model Number 21-7106-24
Device Problems Improper Flow or Infusion (2954); Pressure Problem (3012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/29/2022
Event Type  malfunction  
Event Description
It was reported the pump alarmed high pressure.Patient stated that she changed her cassette at that evening with no issues, but the alarm started going off at around 8 pm.Patient removed the pump from the pouch, checked the pump and it stopped right as she was checking it.The alarm was going off with some but not all her movements.Pharmacist trouble shot with patient while on phone.Patient attached new tubing while on phone with pharmacist and alarm stopped sounding.No patient injury reported.
 
Manufacturer Narrative
A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
Other, other text: no product was returned.We are unable to confirm the reported complaint.If the product is returned, smiths medical will reopen this complaint for further investigation.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.No product was returned; therefore, no visual and functional tests were performed, the reported complaint could not be confirmed, and the root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD EXTENSION SETS
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key16337168
MDR Text Key309176633
Report Number3012307300-2023-01080
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K974013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number21-7106-24
Device Catalogue Number21-7106-24
Device Lot Number4303046
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/12/2023
Initial Date FDA Received02/09/2023
Supplement Dates Manufacturer Received06/29/2023
Supplement Dates FDA Received07/07/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/29/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age55 YR
Patient SexFemale
-
-