Model Number 100129557
Device Problem Connection Problem (2900)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/05/2023
Event Type  malfunction  
Manufacturer Narrative
The results of the investigation were inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information received, the cause of the reported incident could not be determined.
 
Event Description
During an af procedure, it was noted that there were connectiong issues with the ensite dws resulting in multiple restarts which caused delays in treatment.It was noted that the ensite dws was unable to boot up and there was a flashing red light.The dws would close without prompt about 15 seconds after the start button was pressed.The dws was restarted at least five times, but the issue persisted.Cabling was checked, and it was discovered that if the hdmi cable was unplugged, the dws was able to boot.The procedure was completed without adverse patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSITE VELOCITY¿ DWS7 DISPLAY WORKSTATION
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key16338699
MDR Text Key309270641
Report Number2184149-2023-00024
Device Sequence Number1
Product Code DQK
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
K202066
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number100129557
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/07/2023
Initial Date FDA Received02/09/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/08/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-