The returned m540 units along with the masimo and nellcor pods were evaluated with no malfunction found however, the spo2 sensors were found to be at end of life where the sensor no longer lights up as they have ¿timed¿ out but are not faulty.Masimo programs their sensors for a certain amount of uses and once that limit is reached, the sensor will no longer function.The ifu notes, before connecting the sensor to the patient, visually inspect all components for wear or damage.After connection of the sensor observe monitor for any messages if the sensor does not light up.If sensor-led does not light up, then replace sensor.The masimo country manager went onsite where they could not find any environmental issue that attribute to the reported issue.After an extended stay at the customer site, observing the clinical practice showed no malfunction or evidence that may lead to the m540 spo2 readings differing from the abg results.No further information or investigation into the issue was provided from the customer as the hospital is planning to change to philips monitors.Masimo stated they will no longer be investigating the issue as they have exhausted all options with no issues found.Therefore, root cause of the reported issue could not be determined.
|