Model Number DSX500H11C
Device Problems Contamination (1120); Degraded (1153)
Patient Problems Dry Eye(s) (1814); Dyspnea (1816); Headache (1880); Excessive Tear Production (2235); Cough (4457); Unspecified Eye / Vision Problem (4471)
Event Date 10/21/2021
Event Type  malfunction  
Event Description
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging an issue related to a cpap device's sound abatement foam.The patient has alleged to having headaches, watery eyes, dry matter in the corner of the eyes, dry eyes, cough and difficulty breathing.There was no report of patient harm or injury.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Manufacturer Narrative
The manufacturer was previously contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer received information alleging an issue related to a cpap device's sound abatement foam.The patient has alleged to having headaches, watery eyes, dry matter in the corner of the eyes, dry eyes, cough and difficulty breathing.There was no report of serious patient harm or injury.The device was returned to the manufacturer's product investigation laboratory for investigation.An internal visual inspection of the device was completed by the manufacturer and found evidence of water ingress on the blower and blower box.The manufacturer found evidence of dust/dirt contamination in the air inlet, throughout the airpath, on the blower box, and on the motor casing/impellers.The manufacturer found no evidence of sound abatement foam degradation/breakdown.The device's downloaded event log was reviewed by the manufacturer and found no errors logged.The device was applied power and the device operated properly.The manufacturer concludes the contaminates found were consistent with water ingress and dust/dirt contaminant, contamination inconsistent with the sound abatement foam.The manufacturer confirmed there was no evidence of sound abatement foam degradation/breakdown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO CPAP
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key16352170
MDR Text Key309281066
Report Number2518422-2023-04345
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDSX500H11C
Device Catalogue NumberDSX500H11C
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 10/21/2021
Initial Date FDA Received02/10/2023
Supplement Dates Manufacturer Received07/11/2023
Supplement Dates FDA Received07/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/09/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberRES 88058
Patient Sequence Number1
-
-