• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. CONNECTING TUBE; ENDOSCOPE REPROCESSOR ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. CONNECTING TUBE; ENDOSCOPE REPROCESSOR ACCESSORIES Back to Search Results
Model Number MAJ-2113
Device Problem Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/13/2023
Event Type  malfunction  
Manufacturer Narrative
The subject device has not been returned to olympus for evaluation.The investigation is ongoing, and a definitive root cause of the reported event cannot be determined at this time.If additional information becomes available, this report will be supplemented accordingly.
 
Event Description
The customer reported to olympus technical assistance center (tac), during reprocessing, the connecting tube was damaged.There was no harm or user injury reported due to the event.The serial number of the subject device is not provided at this time.This report is related to similar events.(patient identifier: (b)(6)).
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation the specific cause of phenomenon could not be identified.However, it is possible that external force was applied in the wrong direction.The occurrence of the reported problem can be prevented by adhering to the instructions for use (ifu) which states the following: ¿preparation and inspection: move the lock levers of the reprocessor side connector to make sure that they function properly and are not broken.¿.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONNECTING TUBE
Type of Device
ENDOSCOPE REPROCESSOR ACCESSORIES
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16352853
MDR Text Key309313800
Report Number9610595-2023-02242
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170404085
UDI-Public04953170404085
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190969
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 05/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMAJ-2113
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/13/2023
Initial Date FDA Received02/10/2023
Supplement Dates Manufacturer Received04/04/2023
Supplement Dates FDA Received05/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-