• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH T-PAL TRIAL SPACR 10MM X 28MM 7MM HEIGHT; TEMPLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH T-PAL TRIAL SPACR 10MM X 28MM 7MM HEIGHT; TEMPLATE Back to Search Results
Model Number 03.812.307
Device Problems Compatibility Problem (2960); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/18/2023
Event Type  malfunction  
Event Description
Device report from synthes reports an event in norway as follows: it was reported that on january 18, 2023, the trial cage is so worn at the edge facing the inserter that it is not possible to tighten it arrow to arrow.This causes it to turn when inserted before it is suppose to, even if it is in the locked position.The surgery time was extended as a direct result of incident for 10-15 minutes.A new one was used.This report is for one (1) t-pal small trial implant siz 7 non deta.This is report 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Additional narrative: device available for evaluation: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D4: lot number added.H3, h4, h6: the device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: g1: manufacture site updated.H4: manufacture date updated.H3, h4, h6: the product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the t-pal trial spacr 10mm x 28mm 7mm height has signs of normal wear along the surface, the edges of the trial implant are heavily worn and nicked.The observed condition was consistent as an end of life indicator for the device.No other issues were identified.Additionally, the trail implant has a toggling movement and does not longer have a tight fit.After a visual inspection, it was determined that the reusable instrument device was worn from repeated use and servicing; therefore, further investigation for the reported complaint device is not required.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the t-pal trial spacr 10mm x 28mm 7mm height would have contributed to the complained issue.There is no indication that a design or manufacturing issue has caused the complaint condition.It was determined that the reusable instrument was worn from repeated use and servicing.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review part number: 03.812.307 lot number: 58p8685 manufacturing site: haegendorf release to warehouse date: 03.07.2020 a manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T-PAL TRIAL SPACR 10MM X 28MM 7MM HEIGHT
Type of Device
TEMPLATE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK HAEGENDORF
im bifang 6
haegendorf CO 4614
SZ   4614
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key16353142
MDR Text Key309291435
Report Number8030965-2023-01670
Device Sequence Number1
Product Code HWT
UDI-Device Identifier10705034727820
UDI-Public(01)10705034727820
Combination Product (y/n)N
Reporter Country CodeNO
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 02/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number03.812.307
Device Catalogue Number03.812.307
Device Lot Number58P8685
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/19/2023
Initial Date FDA Received02/10/2023
Supplement Dates Manufacturer Received02/17/2023
04/12/2023
Supplement Dates FDA Received03/08/2023
04/12/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/03/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-