It was reported that the patient experienced discomfort in the thoracic area.The physician did not consider the discomfort to be clinically significant.Loss of capture and r-wave amplitude variation were noted on the right ventricular (rv) lead due to dislodgement of the lead.A revision procedure was performed to reposition the rv lead, however, the helix of the rv lead was unable to retract.The rv lead was explanted and replaced to resolve the event.The patient was in stable condition.
|
The reported events were lead dislodgement, loss of capture, sensing r-wave amplitude variation, and helix mechanism issue.As received, a complete lead was returned in one piece for analysis.The reported event of helix mechanism issue was confirmed.Visual inspection found the helix to be partially extended and clogged with blood.X-ray examination found the inner coil to be slightly stretched at the connector region consistent with procedural damage.After cleaning, the helix could be extended and retracted by applying torque to the connector pin.The measured full helix length was within specification.The cause of helix mechanism issue was isolated to the helix being clogged with blood.The reported events of loss of capture and r-wave amplitude variation were not confirmed.Electrical testing did not find any indication of conductor fractures or internal shorts.No anomalies were noted on the lead body at visual inspection.X-ray examination did not find any anomalies except for procedural damage.
|