The field service engineer went on customer site to collect the audit log files.The following functional tests were performed: ¿ logs from the pic ix were obtained.¿ logs have been analyzed by the pse.Results of functional testing indicate there is no malfunction on the pic ix.The audit logs showed there was a red alarm generated for vent fib/tach at 01:36:12 on both the mamh03sp05ipcu and mamhllop06tele hosts (in blue and red).The red alarm kept ringing on host mamh03sp05ipcu until 01:45:55 where it ended.The yellow alarm was generated for the pvc until it was ended at 01:46:37.Overall, there was a red alarm that was sounded until 01:45:55 and then it was turned off, then the yellow alarm went on.After it was turned off (could either have been silenced or that the vtach then did not fit the criteria of having the alarm go red), the user likely saw that yellow alarm relating to the pvc on that host 05ipcu host.The next red alarm occurred on (b)(6), 2023, and 05:01:01 over 3 hours later.Unless there is a reason why this patient should have been red alarmed between 01:45:55 and 05:01:01, the pse does not see any problems with the alarms.Based on the information available and the testing conducted, the device was functioning as intended and there is no malfunction on the device the device was confirmed to be operating per specifications and no failure was identified.
|