• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. OES CYSTONEPHROFIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. OES CYSTONEPHROFIBERSCOPE Back to Search Results
Model Number CYF-5
Device Problems Break (1069); Material Protrusion/Extrusion (2979)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The subject device was received and evaluated.Device evaluation found the device connecting tube is perforated and metal can be seen however, metal protrusion was not observed.Inspection noted there were no metal parts sticking out from the bending rubber.Furthermore, the following defects were identified during device inspection: connecting tube deformed and perforated.Light guide lens cracked.Grip unit discolored.F.Knob (forceps knob) broken.Ks (lever) mouthpiece deformed.Based on device evaluation findings, inspection found that connecting tube is perforated and metal can be seen however, there is no metal protrusion.Multiple follow ups were made to gather additional information regarding the event reported, however, were unsuccessful , no response is received form customer.The replacement of the distal end, the connecting tube unit, the fe-lever, the ks-mouth piece and the k-mount unit are required as a major repair to return the instrument to the original equipment manufacturer (oem) specification.Investigation is ongoing.This report will be supplemented accordingly following investigation.
 
Event Description
Customer sent a repair request reported with an issue of " damaged rubber- visible metal".No harm was reported.No patient harm, no user injury reported.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Three attempts were performed to obtain additional information, but no response was received from the customer.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the reported event is unable to be determined.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES CYSTONEPHROFIBERSCOPE
Type of Device
CYSTONEPHROFIBERSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16382128
MDR Text Key309668017
Report Number3002808148-2023-01454
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170292323
UDI-Public04953170292323
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K221690
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/31/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/20/2023
Initial Date FDA Received02/15/2023
Supplement Dates Manufacturer Received03/16/2023
Supplement Dates FDA Received04/11/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/06/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-