MAUDE Adverse Event Report: PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM Back to Search Results
Model Number AB2000
Device Problems Failure to Prime (1492); Priming Problem (4040)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/18/2023
Event Type  malfunction  
Manufacturer Narrative
Additional manufacturer narrative: device evaluation by manufacturer: device has not yet been returned to manufacturer for evaluation; root cause has not yet been established.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Event Description
A male patient underwent an aquablation procedure for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation (procept) became aware that during the aquablation procedure, the aquabeam handpiece was unable to prime despite multiple troubleshooting steps, including replacing the handpiece with a new handpiece unit, without success.A third handpiece resolved the issue, and the procedure was continued through successful completion.The reported event caused a surgical procedural delay of over 20 minutes.There were no adverse health consequences to the patient due to the procedural delay.
 
Manufacturer Narrative
H.3 evaluation by manufacturer: the aquabeam handpiece was returned for investigation of the reported event.No visual anomalies were observed.Three docking cycles were performed, and after the docking cycles, an attempt to prime the handpiece failed.No fluid was drawn into the inlet tubing, confirming the priming issue observed.Piston assembly was opened, and the position of the inlet valve was slightly adjusted allowing the handpiece to prime.The inlet valve was removed from the piston body inside the pump cartridge to be observed under magnification.No visual defects could be observed.The pump cartridge pulls in the saline into the handpiece through the piston assembly, which moves back and forth rapidly to create vacuum and draw in saline.During this process, the position of the inlet valve keeps changing, which might lead to misalignment.However, root cause of the issue is supplier as the pump cartridge and the inlet valve are supplied through an external supplier.Supplier quality team was informed of this issue.Please see the attached email.A review of the device history record (dhr) for aquabeam robotic system/serial number (b)(6) and the aquabeam handpiece / lot number 22c04014 was performed, which confirmed that there were no nonconformances, failures, discrepancies, or missed steps during the manufacturing process that could be related to the reported event.The review indicated that the system and its associated component met all design and manufacturing specifications when released for distribution.The aquabeam robotic system ifu, ifu0101-00 rev.F, states the following: 8.14 sterile: ensure the aquabeam scope is fully advanced prior to priming.Press the "prime" button on the foot pedal or console and circle priming indicators (100% and 50%) will appear on the cpu.Press the [+] button on either the console or the motorpack to prime the aquabeam handpiece at 100% power level.Continue to press the [+] button until the 100% yellow indicator turns green.Release the [+] button while pressing the prime button until the 50% yellow indicator turns green.The likely root cause of the issue is supplier as the pump cartridge and the inlet valve are supplied through an external supplier.Complaint data is monitored and trended as part of procept's quality management system; shall a trend arise for this failure mode, then further actions will be considered.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUABEAM ROBOTIC SYSTEM
Type of Device
FLUID JET REMOVAL SYSTEM
Manufacturer (Section D)
PROCEPT BIOROBOTICS CORPORATION
900 island drive
suite 101
redwood city CA 94065 1494
Manufacturer (Section G)
PROCEPT BIOROBOTICS CORPORATION
900 island drive
suite 101
redwood city CA 94065 1494
Manufacturer Contact
doria esquivel
900 island drive
suite 101
redwood city, CA 94065-1494
MDR Report Key16387576
MDR Text Key309743981
Report Number3012977056-2023-00019
Device Sequence Number1
Product Code PZP
UDI-Device IdentifierB614AB20001
UDI-Public+B614AB20001/$$722C00188/16D20220204E
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
DEN170024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAB2000
Device Catalogue NumberAB2000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/18/2023
Initial Date FDA Received02/16/2023
Supplement Dates Manufacturer Received03/23/2023
Supplement Dates FDA Received03/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/04/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexMale
-
-