A male patient underwent an aquablation procedure for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation (procept) became aware that during the aquablation procedure, the aquabeam handpiece was unable to prime despite multiple troubleshooting steps, including replacing the handpiece with a new handpiece unit, without success.A third handpiece resolved the issue, and the procedure was continued through successful completion.The reported event caused a surgical procedural delay of over 20 minutes.There were no adverse health consequences to the patient due to the procedural delay.
|
H.3 evaluation by manufacturer: the aquabeam handpiece was returned for investigation of the reported event.No visual anomalies were observed.Three docking cycles were performed, and after the docking cycles, an attempt to prime the handpiece failed.No fluid was drawn into the inlet tubing, confirming the priming issue observed.Piston assembly was opened, and the position of the inlet valve was slightly adjusted allowing the handpiece to prime.The inlet valve was removed from the piston body inside the pump cartridge to be observed under magnification.No visual defects could be observed.The pump cartridge pulls in the saline into the handpiece through the piston assembly, which moves back and forth rapidly to create vacuum and draw in saline.During this process, the position of the inlet valve keeps changing, which might lead to misalignment.However, root cause of the issue is supplier as the pump cartridge and the inlet valve are supplied through an external supplier.Supplier quality team was informed of this issue.Please see the attached email.A review of the device history record (dhr) for aquabeam robotic system/serial number (b)(6) and the aquabeam handpiece / lot number 22c04014 was performed, which confirmed that there were no nonconformances, failures, discrepancies, or missed steps during the manufacturing process that could be related to the reported event.The review indicated that the system and its associated component met all design and manufacturing specifications when released for distribution.The aquabeam robotic system ifu, ifu0101-00 rev.F, states the following: 8.14 sterile: ensure the aquabeam scope is fully advanced prior to priming.Press the "prime" button on the foot pedal or console and circle priming indicators (100% and 50%) will appear on the cpu.Press the [+] button on either the console or the motorpack to prime the aquabeam handpiece at 100% power level.Continue to press the [+] button until the 100% yellow indicator turns green.Release the [+] button while pressing the prime button until the 50% yellow indicator turns green.The likely root cause of the issue is supplier as the pump cartridge and the inlet valve are supplied through an external supplier.Complaint data is monitored and trended as part of procept's quality management system; shall a trend arise for this failure mode, then further actions will be considered.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
|