• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL REGENT HEART VALVE WITH FLEX CUFF; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL REGENT HEART VALVE WITH FLEX CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Model Number AGFN-756
Device Problems Incomplete Coaptation (2507); Patient Device Interaction Problem (4001)
Patient Problems Arrhythmia (1721); Thrombosis/Thrombus (4440)
Event Date 01/02/2023
Event Type  Injury  
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that on (b)(6) 2022, a 21mm sjm regent heart valve w/flex cuff was successfully implanted.On an unknown date after the valve was implanted, the patient developed atrial arrhythmia.Thrombus formation was seen in the atria and led to fixation of one of the valve lobes.On (b)(6) 2023, re-operation was performed due to valve opening failure.During this procedure, it was seen that a ring-shaped thrombus had formed on the left atrial wall around the annulus and had spread to the mechanical valve.The left atrial side of the valve was covered by a thin thrombus like material.The valve was explanted.No additional information was provided.
 
Event Description
Subsequent to the previously filed report, additional information was received: it was noted that the valve remained open after thrombus occurred.
 
Manufacturer Narrative
Explant due to the valve not opening properly and a thrombus was reported.The investigation found that the mechanical leaflets opened and closed completely and were freely mobile.No material was seen within pivot recesses or adherent to mechanical leaflets.There was focal adherent dark red-brown material on the sewing cuff.A piece of material removed from the sewing cuff consists of fibrin and clotted blood.No inflammation or significant calcifications were present.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and that the product met all specifications.The cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REGENT HEART VALVE WITH FLEX CUFF
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key16390261
MDR Text Key309750029
Report Number2135147-2023-00554
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier05414734005975
UDI-Public05414734005975
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAGFN-756
Device Catalogue Number21AGFN-756
Device Lot NumberC00005363
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/25/2023
Initial Date FDA Received02/16/2023
Supplement Dates Manufacturer Received04/03/2023
Supplement Dates FDA Received04/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/25/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
-
-