Catalog Number 04625358019
Device Problem High Test Results (2457)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/17/2023
Event Type  malfunction  
Manufacturer Narrative
Occupation is patient/consumer.The strips were requested for investigation.The product has not been received at this time.If the product is returned in the future, a follow up report will be submitted.On a regular basis, coaguchek strips of lots currently valid in the market are tested as part of routine retention testing and results have passed the internal inspection.Per product labeling: coaguchek uses human recombinant thromboplastin.Therefore, the comparability to other human recombinant thromboplastins is best, whereas higher deviations can occur with other thromboplastins.However, those higher differences between thromboplastins of different (rabbit, bovine based) origin are not a coaguchek specific issue.Similar differences can be observed when a human recombinant thromboplastin-based laboratory method is compared against several other (rabbit, bovine-based) laboratory methods.
 
Event Description
We received an allegation of questionable inr results for 1 patient tested with coaguchek xs meter serial number (b)(4) compared to an unknown laboratory method.On (b)(6) 2023 the patient had a laboratory result of 1.25 inr while at 4:48 p.M.The meter result was 2.0 inr.The time difference bewteen the measurements was 1 hour.The patient's therapeutic range was 2.5-3.0 inr.The patient's interval of testing is once a week.
 
Manufacturer Narrative
The patient code and the evaluation conclusion code were updated.Per product labeling: the action of oral anticoagulants (coumarin derivatives) can be increased or weakened when other medication is taken simultaneously (e.G.Antibiotics).This, in turn, can also lead to either an increase or a decrease in prothrombin time (inr).If other medication is taken, it is recommended that the prothrombin time be checked more frequently and that the anticoagulant dose be subsequently adjusted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS PT TEST
Type of Device
PROTHROMBIN TIME TEST STRIPS
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key16395601
MDR Text Key309751307
Report Number1823260-2023-00487
Device Sequence Number1
Product Code GJS
Combination Product (y/n)N
Reporter Country CodeTU
PMA/PMN Number
K062925
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/31/2024
Device Catalogue Number04625358019
Device Lot Number63312513
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/20/2023
Initial Date FDA Received02/17/2023
Supplement Dates Manufacturer Received03/03/2023
Supplement Dates FDA Received03/30/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ANTIBIOTICS; VITAMIN D
-
-