MAUDE Adverse Event Report: DATEX-OHMEDA, INC. AVANCE; ANESTHESIA GAS MACHINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATEX-OHMEDA, INC. AVANCE; ANESTHESIA GAS MACHINE Back to Search Results
Device Problem Insufficient Information (3190)
Patient Problem Cardiac Arrest (1762)
Event Date 01/18/2023
Event Type  Injury  
Manufacturer Narrative
Ge healthcare's investigation into the reported occurrence is ongoing.A follow-up report will be issued when the investigation has been completed.Customer reported no patient information available.Legal manufacturer: hcs madison - 3030 ohmeda dr, usa madison, wi 53718.Device evaluation anticipated, but not yet begun.
 
Event Description
The hospital reported a patient was connected to an avance when the patient coded.The unit was replaced, and the case was resumed.There was no patient injury.Ge healthcare will submit a follow-up report when the investigation has been completed.
 
Manufacturer Narrative
Ge healthcare engineering investigated the reported event.The root cause is unknown though it is possible that there was a leak or occlusion in the patient circuit.There was no allegation from the customer that the device contributed to the patient coding.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVANCE
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
anthony amenson
3030 ohmeda drive
madison, WI 53718
MDR Report Key16395625
MDR Text Key309737772
Report Number2112667-2023-00764
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172045
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 04/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/19/2023
Initial Date FDA Received02/17/2023
Supplement Dates Manufacturer Received03/24/2023
Supplement Dates FDA Received04/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/11/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-