Catalog Number 829555
Device Problems Product Quality Problem (1506); Manufacturing, Packaging or Shipping Problem (2975)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/06/2023
Event Type  malfunction  
Event Description
End user reports that one syringe from a box of 100 of syringes of item number 829555 lot number 57091b expiration date 12/02/2026 felt different while injecting into the skin.Upon the end user's own investigation with a magnifying glass, the end of the cannula was burred.
 
Manufacturer Narrative
Initial trend analysis for lot 57091b was conducted, no malfunctions were found.This is the only complaint for lot 57091b.Further investigation will be conducted to determine the root cause of complaint.
 
Event Description
End user reports that one syringe from a box of (b)(4) of syringes of item number 829555 lot number 57091b expiration date 12/02/2026 felt different while injecting into the skin.Upon the end user's own investigation with a magnifying glass, the end of the cannula was burred.
 
Manufacturer Narrative
No device was returned for testing.Retained lot number 57091b were investigated for needle sharpness test and visual inspection.All 50 retained lot samples passed investigation tests, no abnormalities found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYTOUCH
Type of Device
SYRINGE
Manufacturer (Section D)
MHC MEDICAL PRODUCTS, LLC
8695 seward road
fairfield OH 45011
Manufacturer (Section G)
MHC MEDICAL PRODUCTS, LLC
8695 seward road
fairfield OH 45011
Manufacturer Contact
troy smith
8695 seward road
fairfield, OH 45011
MDR Report Key16396686
MDR Text Key310152189
Report Number3005798905-2023-03101
Device Sequence Number1
Product Code FMF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102178
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Remedial Action Replace
Type of Report Initial,Followup
Report Date 02/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number829555
Device Lot Number57091B
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/06/2023
Initial Date FDA Received02/17/2023
Supplement Dates Manufacturer Received02/06/2023
Supplement Dates FDA Received03/22/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-