Complainant alleged that while attempting to defibrillate an 88-year-old male patient, a spark was seen from the sternum electrode pad during the fourth discharge.After removing the electrode pads, second degree burns were found on the patient's skin.Complainant indicated that the clinician obtained another set of electrode pads to continue treating the patient.
|
The onestep cpr pads were returned to zoll medical corporation for evaluation.The customer's report was not replicated or confirmed.The onestep cpr pads were put through extensive testing and successfully acquired a valid impedance and discharged appropriately.No anomalies were observed during visual inspection.Replacement pads were sent to the customer.No trend is associated with reports of this type.It is important to note that during defibrillation therapy, patients can experience burns and redness due to a high measured patient impedance.This could be for a variety of reasons including, but not limited to, skin preparation, electrode application, or poor coupling.
|