Philips received a complaint on the intellivue patient monitor mx850, indicating displays a speaker malfunction message.No harm to the patient.The field service engineer performed an analysis on the equipment and did not confirm the reported failure.The equipment shows a speaker malfunction message and the sound was coming from the mx800.After restarting the device, it performed as expected and the equipment was released for use.The device was confirmed to be operating per specifications and no failure was identified.The investigation concludes that no further action is required at this time.If additional information is received, the claim file will be reopened.
|