H4: the lot was manufactured from march 22, 2022 - march 30, 2022.The device was received for evaluation.A visual inspection with the naked eye and under the microscope was performed which showed no evidence of a cut on the fill port.The fill port was found to be in normal condition.The thin line that is on the stressmember flange near the fill port is not a cut.The thin line is a mold design of the stressmember flange (a design of the product).Based on the evaluation result, the infusor sample was determined to be conforming product.Therefore, the reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|