• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION INFUSOR; PUMP, INFUSION, ELASTOMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION INFUSOR; PUMP, INFUSION, ELASTOMERIC Back to Search Results
Catalog Number J2C1711
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/27/2023
Event Type  malfunction  
Manufacturer Narrative
The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a small volume infusor had a cut on the filling port.This issue was discovered during filling.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H4: the lot was manufactured from march 22, 2022 - march 30, 2022.The device was received for evaluation.A visual inspection with the naked eye and under the microscope was performed which showed no evidence of a cut on the fill port.The fill port was found to be in normal condition.The thin line that is on the stressmember flange near the fill port is not a cut.The thin line is a mold design of the stressmember flange (a design of the product).Based on the evaluation result, the infusor sample was determined to be conforming product.Therefore, the reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key16404758
MDR Text Key310071881
Report Number1416980-2023-00573
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412092706
UDI-Public(01)00085412092706
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberJ2C1711
Device Lot Number22C032
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2023
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/27/2023
Initial Date FDA Received02/20/2023
Supplement Dates Manufacturer Received03/13/2023
Supplement Dates FDA Received03/24/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-