ZOLL MEDICAL CORPORATION AED3; DEFIBRILLATOR Back to Search Results
Model Number AED3
Device Problem Incorrect Interpretation of Signal (1543)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/20/2023
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event Description
Complainant alleged that while attempting to monitor a male patient (age unknown), the device issued a "shock advised" prompt for a heart rhythm they believe was non-shockable.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
 
Manufacturer Narrative
The device was returned to zoll medical corporation for evaluation.The customer's report was attributed to the incorrect use of the device by the user.The device was used on a conscious patient.The aed 3 administrator's guide and operator's guide both explain the contraindications for use as: never use the zoll aed 3 for defibrillation when the patient is conscious, is breathing, or has a detectable pulse or other sign of circulation.The device passed all testing, was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AED3
Type of Device
DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key16405189
MDR Text Key309985095
Report Number1220908-2023-00508
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00847946020071
UDI-Public00847946020071
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberAED3
Device Catalogue NumberAED3
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/24/2023
Initial Date FDA Received02/20/2023
Supplement Dates Manufacturer Received01/24/2023
Supplement Dates FDA Received04/12/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexMale
-
-