Model Number E411 RACK |
Device Problems
Incorrect, Inadequate or Imprecise Result or Readings (1535); High Test Results (2457); Low Readings (2460)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/25/2023 |
Event Type
malfunction
|
Event Description
|
The initial reporter complained of qc issues after replacing the probnp reagent on a cobas e 411 immunoassay analyzer.After noticing the qc issue, the customer repeated patient samples and there was a difference in results.The following are examples of discrepant results for 5 patient samples.Note: the unit of measure used by the customer is not clear.This information has been requested but has not been provided.Patient 1 initial result was 129.The repeat result was 87.Patient 2 initial result was 631.The repeat was 832.Patient 3 initial result was 2615.The repeat was 3436.Patient 4 initial result was 642.The repeat was 859.Patient 5 initial result was 449.The repeat was 585.The questionable results were reported outside of the laboratory.The probnp reagent lot number and expiration date were not provided.
|
|
Manufacturer Narrative
|
The suspect medical device was updated.Applicable fields of sections d and g were updated.The e411 analyzer serial number was (b)(6).
|
|
Manufacturer Narrative
|
Calibration was acceptable.Third-party qc was acceptable.The measuring cell was last replaced on 15-jun-2022.When a new reagent kit with the same lot number was loaded onto the instrument, qc was performed.Qc results indicated a new calibration was required.After calibration, qc was within range.The initial patient results were obtained before the new calibration was performed; the repeat results were obtained after recalibration.The investigation did not identify a product problem.The cause of the event could not be determined.
|
|
Search Alerts/Recalls
|