Information received from healthcare provider via manufacturer representative regarding an event happened during intra-op of the reported product.The pre-op diagnosis was l4/5 hernia.It was reported that, the flexible arm tightening felt loose during the med l4/5 procedure, but the operation was performed as is.There was no delay in overall procedural time reported.No patient symptoms or complications reported.No additional treatment or surgery performed as a result.No further complications reported.
|
Product analysis of part# 9560524, lot# my21k001.Service of reported product confirmed, the screw thread of the handle screw was deformed.Root cause was unknown.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|