CARDINAL HEALTH ENTRFLX 10FR 43IN W STY YPORT; TUBES, GASTROINTESTINAL (AND ACCESSORIES)
|
Back to Search Results |
|
Model Number 8884721088E |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/17/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
|
Event Description
|
Per customer, the nurse stated that she was unscrewing the feeding from the feeding tube and the enfit end separated from the feeding tube.There was no harm reported.
|
|
Manufacturer Narrative
|
Since a lot number was not provided, the device history record review could not be performed.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.One sample was received for analysis.Upon analysis, disassembly issue was confirmed.An exact root cause of the reported issue could not be determined.At this time, a corrective and preventive action is not deemed necessary.We will keep monitoring the process for any adverse trends that require immediate attention.This complaint will be used for qa tracking and trending purposes.
|
|
Manufacturer Narrative
|
Since a lot number was not provided, the device history record review could not be performed.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.A picture and one sample was received for analysis.Upon analysis, disassembly issue was confirmed.An exact root cause of the reported issue could not be determined.At this time, a corrective and preventive action is not deemed necessary.We will keep monitoring the process for any adverse trends that require immediate attention.This complaint will be used for qa tracking and trending purposes.
|
|