MAUDE Adverse Event Report: CARDINAL HEALTH ENTRFLX 10FR 43IN W STY YPORT; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH ENTRFLX 10FR 43IN W STY YPORT; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 8884721088E
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/17/2023
Event Type  malfunction  
Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event Description
Per customer, the nurse stated that she was unscrewing the feeding from the feeding tube and the enfit end separated from the feeding tube.There was no harm reported.
 
Manufacturer Narrative
Since a lot number was not provided, the device history record review could not be performed.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.One sample was received for analysis.Upon analysis, disassembly issue was confirmed.An exact root cause of the reported issue could not be determined.At this time, a corrective and preventive action is not deemed necessary.We will keep monitoring the process for any adverse trends that require immediate attention.This complaint will be used for qa tracking and trending purposes.
 
Manufacturer Narrative
Since a lot number was not provided, the device history record review could not be performed.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.A picture and one sample was received for analysis.Upon analysis, disassembly issue was confirmed.An exact root cause of the reported issue could not be determined.At this time, a corrective and preventive action is not deemed necessary.We will keep monitoring the process for any adverse trends that require immediate attention.This complaint will be used for qa tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTRFLX 10FR 43IN W STY YPORT
Type of Device
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key16409575
MDR Text Key309949427
Report Number9612030-2023-03555
Device Sequence Number1
Product Code KNT
UDI-Device Identifier10884521582996
UDI-Public10884521582996
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date 05/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8884721088E
Device Catalogue Number8884721088E
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/20/2023
Initial Date FDA Received02/21/2023
Supplement Dates Manufacturer Received02/20/2023
02/20/2023
Supplement Dates FDA Received05/02/2023
05/02/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage A
Patient Sequence Number1
-
-