|
Model Number 45031 |
Device Problems
Display or Visual Feedback Problem (1184); Pumping Problem (3016)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/04/2023 |
Event Type
malfunction
|
Event Description
|
Reportable based on device analysis completed on 02-feb-2023.It was reported that an error occurred.The target location was located in the iliac vein of the left lower limb.An angiojet solent omni was used for a thrombectomy procedure.During the procedure, it was noted that the system alerted an error, and the sac inlet was not working properly.The procedure was completed with another of the same device.There were no patient complications, and the patient was stable.However, returned device analysis revealed a broken hypotube.
|
|
Manufacturer Narrative
|
(b)(6).Device evaluated by mfr: the product was returned to boston scientific for analysis.Returned product consisted of an angiojet solent omni catheter.The pump assembly, effluent/supply line, shaft, tip, and spike line were visually examined for damage or any irregularities.The catheter shaft showed 2 kinks located.5cm and 83.5cm from the tip.Functional testing was attempted; however, the device would not prime.During analysis it was noticed that the hypotube was broken.5cm from the tip at a severely kinked area.X-ray analysis confirmed the break.The device could not be functionally tested due to the extreme damage on the device.No pressure reading could be reported.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.
|
|
Search Alerts/Recalls
|
|
|