• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID W/ E/F PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID W/ E/F PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-55
Device Problem Component Incompatible (1108)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/13/2023
Event Type  malfunction  
Event Description
It was reported that, the cardiosave intra-aortic balloon pump (iabp) unit has a bleed purge problem.
 
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that before patient use, the cardiosave intra-aortic balloon pump (iabp) unit has a bleed purge problem.There was no patient harm reported.
 
Event Description
It was reported that during preparation, before patient use, the cardiosave intra-aortic balloon pump (iabp) unit had a bleed purge problem.It was just before the ball was to be put in place that they had a bleeding problem.There was no patient harm reported.
 
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to evaluate the iabp unit and the fse noticed that there was water between the safety disk and the pim, so the fse did a water purge and the test passed.The fse did not see any problem when checking the machine.The fse found a problem with the printer.Estimate in progress.The fse changed the exec card because it had not be changed last year during preventive maintenance.Checking of the device according to the manufacturer's procedure.If information is received for the printer, we will reopen and update the complaint.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID W/ E/F PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key16416681
MDR Text Key310069357
Report Number2249723-2023-01221
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108414
UDI-Public10607567108414
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup,Followup
Report Date 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0998-00-0800-55
Device Catalogue Number0998-00-0800-55
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/13/2023
Initial Date FDA Received02/22/2023
Supplement Dates Manufacturer Received02/23/2023
03/02/2023
04/16/2024
Supplement Dates FDA Received03/01/2023
03/06/2023
04/17/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/10/2012
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
UNKNOWN.
-
-