Device Problem Expulsion (2933)
Patient Problem Insufficient Information (4580)
Event Date 01/04/2023
Event Type  Injury  
Manufacturer Narrative
It was reported that a patient was implanted with a with a prodisc l implant at l5-s1, sometime in the summer of 2022.Approximately 6 months later, it was found that the poly inlay had slid anteriorly and the inlay was expulsed.A revision surgery was completed on (b)(6) 2023 to remove the old poly inlay and seat a new inlay within the pdl implant.A dhr review could not be completed as the part number and lot number were not provided and could not be determined from the information received as part of the investigation.Complaint trending found that the rate of complaints is within the poc level defined within the risk documentation.A review of the risk documentation found that the hazard associated with this complaint is identified and mitigated to a level where the benefits outweigh the risks.Device evaluation could not be completed because although the poly inlay was returned it was very damaged from the removal process.It was confirmed that a revision took place due to poly inlay expulsion.No other anomalies associated with the complaint were found during the investigation.This report is for 3 of 3 devices involved in this event.
 
Event Description
Patient was implanted with a with a prodisc l implant at l5-s1, sometime in the summer of 2022.Approximately 6 months later, it was found that the poly inlay had slid anteriorly and the inlay was expulsed.A revision surgery was completed on (b)(6) 2023 to remove the old poly inlay and seat a new inlay within the pdl implant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRODISC L INFERIOR PLATE, LARGE
Type of Device
PROSTHESIS, INTERVERTEBRAL DISC
Manufacturer (Section D)
CENTINEL SPINE, LLC
900 airport road, suite 3b
west chester PA 19380
Manufacturer Contact
erin ward
900 airport road
suite 3b
west chester, PA 19380
4848878894
MDR Report Key16426512
MDR Text Key310049403
Report Number3007494564-2023-00007
Device Sequence Number1
Product Code MJO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P050010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/26/2023
Initial Date FDA Received02/23/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age45 YR
Patient SexMale
-
-