MAUDE Adverse Event Report: SYNTHES GMBH RETENT PIN SCRDRIVER QC -HEX 12/ XL25; SCREWDRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH RETENT PIN SCRDRIVER QC -HEX 12/ XL25; SCREWDRIVER Back to Search Results
Catalog Number 03.045.006
Device Problem Material Twisted/Bent (2981)
Patient Problem Unspecified Tissue Injury (4559)
Event Date 02/02/2023
Event Type  Injury  
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device available for evaluation: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on (b)(6) 2023, the threads of the long retention pin stripped while inserting a 5.0x40mm screw in the proximal femur of an rfna.There was surgical delay of 5 minutes.The surgeon extended the incision and used additional fluoroscopy to retrieve the screw.There were no patient consequences.This report involves one retention pin for screwdriver w/ q c -hex 12mm/ xl25.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.Part number: 03.045.006, lot number: 82p0017, manufacturing site: haegendorf, release to warehouse date: 26.01.2021.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the threads of the distal tip of the scrdriver qc -hex 12/ xl25 were observed stripped and dented.No other issue was identify.A dimensional inspection for the scrdriver qc -hex 12/ xl25 was unable to be performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the scrdriver qc -hex 12/ xl25 would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RETENT PIN SCRDRIVER QC -HEX 12/ XL25
Type of Device
SCREWDRIVER
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK HÄGENDORF (CH)
im bifang 6
haegendorf CO 4614
SZ   4614
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key16429541
MDR Text Key310092305
Report Number8030965-2023-02233
Device Sequence Number1
Product Code HXX
UDI-Device Identifier07612334167835
UDI-Public(01)07612334167835
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03.045.006
Device Lot Number82P0017
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/18/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/02/2023
Initial Date FDA Received02/23/2023
Supplement Dates Manufacturer Received03/01/2023
Supplement Dates FDA Received03/16/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/26/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
LOCKING SCREW FOR IM NAIL 5/ 40/ XL25
Patient Outcome(s) Required Intervention;
Patient SexMale
-
-