Model Number 10033 |
Device Problems
Device Emits Odor (1425); Environmental Particulates (2930)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/31/2023 |
Event Type
malfunction
|
Event Description
|
A customer reported an event of an odor/smell and mist/haze emitting from the sterrad® nx sterilizer.There was no report of any injuries or human reactions.The customer was instructed to power off the unit and discontinue use.An asp field service engineer was dispatched to assess the unit onsite.This event is being reported as a malfunction report subsequent to a previous serious injury.
|
|
Manufacturer Narrative
|
A field service engineer was dispatched to the customer site.The adapter converter, catalytic converter, oil mist filter, and vacuum pump oil were replaced to resolve the odor/smells and mist/haze issues.Unit meets specifications and was returned to service.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by advanced sterilization products, or its employees that the report constitutes an admission that the product, advanced sterilization products, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Asp complaint ref #: cmp-000118099.
|
|
Manufacturer Narrative
|
H3: asp investigation summary: the investigation included a review of the device history record (dhr), trending analysis of the odor/smells and mist/haze issue, and system risk analysis (sra).¿ the device history record (dhr) was reviewed and no issues relating to the failure mode were noted.The involved unit met manufacturer specifications at the time of release.¿ trending analysis of the odor/smells and mist/haze issues for the sterrad® nx unit was reviewed for the prior six months from the event date, and no significant trend was observed.¿ review of risk documentation shows the risk for exposure to toxic or corrosive material to be "low." no parts were available for further analysis.The assignable cause of the odor/smell and mist/haze is potentially due to the adapter converter, catalytic converter, oil mist filter, and vacuum pump oil.The asp field service engineer replaced the parts and confirmed the sterrad® nx met functional specifications after service.The issues were resolved at the customer facility.Asp will continue to track and trend this issue.Asp complaint ref #: (b)(4).
|
|
Search Alerts/Recalls
|
|