THORATEC SWITZERLAND GMBH CENTRIMAG 2ND GENERATION PRIMARY CONSOLE; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 201-30300
Device Problems No Display/Image (1183); Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/09/2023
Event Type  malfunction  
Manufacturer Narrative
Manufacturer's investigation conclusion: the reported event of unknown audio alert without any error messages and the reported event of the display going black 10 ¿ 15 seconds following the alert was not confirmed.The centrimag 2nd generation primary console (serial #: (b)(4) was not returned for analysis.Additional provided information communicated on 16feb2023 stated that only the console was exchanged.No product will be returned for evaluation.The hospital left the console running for 24 hours without any issues.The reported event was unable to be reproduced again.The root cause for the reported event was unable to be conclusively determined through this analysis.The device history records were reviewed for the centrimag 2nd generation primary console (serial #: (b)(4) and the console was found to pass all manufacturing and qa specifications.The 2nd generation centrimag system operating manual section 4 entitled "warnings & precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction".The 2nd generation centrimag system operating manual entitled "emergency and troubleshooting" states that "the recommended practice whenever there is a 2nd generation centrimag primary console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the backup motor and console to continue patient support.Do not exchange individual motors or individual consoles during patient support".The 2nd generation centrimag system operating manual ¿appendix i ¿ console alarms and alerts¿ table 16 details how to properly interpret and troubleshoot all system alarms.No further information was provided.The manufacturer is closing the file on this event.
 
Event Description
It was reported that the centrimag 2nd generation primary console had an audio alert without any error messages on (b)(6) 2023.The display also went black for 10-15 seconds.The patient had used the centrimag since their implant on (b)(6) 2023.When the issue with the console occurred on (b)(6) 2023, the device was exchanged.There were no patient consequences.The exchanged console was then connected to a motor without a patient and the centrimag was functioning properly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIMAG 2ND GENERATION PRIMARY CONSOLE
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ  CH-8005
Manufacturer (Section G)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ   CH-8005
Manufacturer Contact
bob fryc
6035 stoneridge drive
pleasanton, CA 94588
7818528204
MDR Report Key16433476
MDR Text Key310420754
Report Number3003306248-2023-00752
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K131179
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number201-30300
Device Lot NumberL06525
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/09/2023
Initial Date FDA Received02/23/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/11/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-