Model Number 470056-05
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/28/2022
Event Type  malfunction  
Event Description
It was reported that the 0-degree endoscope plus had a cut in the cord.There was no report of patient involvement.
 
Manufacturer Narrative
Based on the claim against the product by the customer noting cut in cord, an investigation was completed to determine the cause of this reported event.Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis.The 0-degree endoscope plus was analyzed and found to have cable integrity damage at zone c and laser markings on housing damage/markings on housing nose/shell housing.The 0-degree endoscope plus had an error 45312 in the right eye.The voltage test and vch current test failed.The endoscope had missing endoscope adapter (aea) retaining ring or screws, discolored housing, and visible light through cable jacket.This complaint is a reportable malfunction event due to the following conclusion: the xi endoscope was found with a missing camera instrument adapter (aea) component.A missing retaining ring results in poor camera control, which could result in unintuitive motion and subsequent tissue damage.While there was no harm or injury to the patient, the reported failure mode could likely cause or contribute to an adverse event if it were to recur.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NONE
Type of Device
ENDOSCOPE
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key16440336
MDR Text Key310286760
Report Number2955842-2023-10898
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier00886874116562
UDI-Public(01)00886874116562
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191736
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Biomedical Engineer
Remedial Action Other
Type of Report Initial
Report Date 01/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number470056-05
Device Catalogue Number470056
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/31/2023
Initial Date FDA Received02/24/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES.
-
-