MEDTRONIC MED REL MEDTRONIC PUERTO RICO ADAPTA; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number ADDR01
Device Problems Pacing Problem (1439); Battery Problem: High Impedance (2947); Reset Problem (3019)
Patient Problem Bradycardia (1751)
Event Date 02/03/2023
Event Type  malfunction  
Event Description
It was reported that the patient experienced pacing below the set rate.It was further reported that the implantable pulse generator (ipg) exhibited a power on reset and had increased battery impedance.It was also reported that the right ventricular (rv) lead exhibited no capture.The ipg and rv lead remain in use.No further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADAPTA
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key16441912
MDR Text Key310255714
Report Number3004209178-2023-02618
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier00613994222046
UDI-Public00613994222046
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P980035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/28/2010
Device Model NumberADDR01
Device Catalogue NumberADDR01
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/03/2023
Initial Date FDA Received02/27/2023
Date Device Manufactured06/03/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
5054-52 LEAD, 5554-45 LEAD.
Patient Age95 YR
Patient SexFemale
-
-