Model Number D-AVHD-DF16
Device Problems Display or Visual Feedback Problem (1184); Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/21/2023
Event Type  malfunction  
Event Description
During the atrial fibrillation procedure, the catheter was removed from the sheath and damage was noted on the device.It was noted the a1 electrode of the advisor hd grid catheter was not displayed correctly.The shaft appeared to be torn and an electrode appeared to be shifted.The procedure was completed successfully with no adverse patient consequences.
 
Manufacturer Narrative
One bi-directional, curve d-f, sensor enabled, advisor hd grid mapping catheter was received for evaluation.Electrodes 1-2 read as open circuits.Conductor wires 1-2 had been fractured at the ring joint, consistent with the reported display issue and noise.In addition, the electrodes were displaced and the pellethane tubing was torn and corrugated.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The root cause of the fractured conductor wires, displaced electrodes, and torn and corrugated pellethane tubing remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVISOR¿ HD GRID MAPPING CATHETER, SENSOR ENABLED¿
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key16448861
MDR Text Key310341482
Report Number3005334138-2023-00076
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K172393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberD-AVHD-DF16
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/21/2023
Initial Date FDA Received02/27/2023
Supplement Dates Manufacturer Received05/23/2023
Supplement Dates FDA Received05/26/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-