STRYKER ORTHOPAEDICS-MAHWAH 22.2MM STD LFIT V40 HEAD; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
|
Back to Search Results |
|
Catalog Number 6260-9-122 |
Device Problem
Material Erosion (1214)
|
Patient Problems
Foreign Body Reaction (1868); Metal Related Pathology (4530)
|
Event Date 02/02/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Reported event: an event regarding abnormal ion level involving a metal head was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: could not be performed as the device lot details were not provided.Complaint history review: could not be performed as the device lot details were not provided.Conclusions: the event could not be confirmed as insufficient information was provided.Further information such as device identification and return, pre- and post-operative x-rays, primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
|
Event Description
|
It was reported by the patient's attorney as a result of a legal claim that allegedly on or about (b)(6) 2019, the patient underwent a right total hip replacement with stryker components.It is further alleged that three months after the right hip replacement, the plaintiff began to experience clicking of the joint, along with pain in her hip, groin and buttock area.Blood test revealed elevated chromium (1.0) and cobalt (7.3) levels as well as elevated esr (9) and crp (2.8) levels.Allegedly the patient was revised on an unknown date.
|
|
Manufacturer Narrative
|
An event regarding wear (metallosis) and abnormal ion level involving a metal head was reported.Wear was confirmed through clinician review.Abnormal ion level was not confirmed.Method & results: -product evaluation and results: visual inspection, material analysis, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: a review of medical records with a clinical consultant indicated "as described above, this product inquiry concerns a patient who underwent a primary total hip arthroplasty with a cobalt chrome head, mdm articulation, with an accolade ii femoral stem.The patient developed pain and within about two years was diagnosed with metallosis and required revision which was performed with a competitor product.I can confirm that the patient had the primary and revision procedure since i was able to review the operation report.The root cause of this event cannot be determined with certainty.The causes of metallosis are multi-factorial including surgical technique, specifically in the preparation or lack thereof, of the femoral head or acetabular liner prior to insertion on the trunnion and into the acetabular shell.Other factors include the patient's activity level and bmi, as well as implant factors." -product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: a review of medical records with a clinical consultant indicated "as described above, this product inquiry concerns a patient who underwent a primary total hip arthroplasty with a cobalt chrome head, mdm articulation, with an accolade ii femoral stem.The patient developed pain and within about two years was diagnosed with metallosis and required revision which was performed with a competitor product.I can confirm that the patient had the primary and revision procedure since i was able to review the operation report.The root cause of this event cannot be determined with certainty.The causes of metallosis are multi-factorial including surgical technique, specifically in the preparation or lack thereof, of the femoral head or acetabular liner prior to insertion on the trunnion and into the acetabular shell.Other factors include the patient's activity level and bmi, as well as implant factors." no further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|
|
Event Description
|
It was reported by the patient's attorney as a result of a legal claim that allegedly on or about (b)(6) 2019, the patient underwent a right total hip replacement with stryker components.It is further alleged that three months after the right hip replacement, the plaintiff began to experience clicking of the joint, along with pain in her hip, groin and buttock area.Blood test revealed elevated chromium (1.0) and cobalt (7.3) levels as well as elevated esr (9) and crp (2.8) levels.Allegedly the patient was revised on an unknown date.
|
|
Search Alerts/Recalls
|
|
|