• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US DXTEND SCREW LOCK D4.5X36MM; DELTA XTEND IMPLANTS : SCREWS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US DXTEND SCREW LOCK D4.5X36MM; DELTA XTEND IMPLANTS : SCREWS Back to Search Results
Model Number 130790036
Device Problem Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/27/2022
Event Type  malfunction  
Event Description
It was reported that during a shoulder prosthesis procedure (delta xtend), 1 screw broke and the second one got bent.They had to complete the procedure with another samelike screw.15 minutes of surgical delay.No patient consequence.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this report was returned for analysis.Visual analysis revealed that the prongs of the dxtend screw lock d4.5x36mm were found twisted, additionally three prongs were found broken, therefore the report allegation can be confirmed.The broken fragment was not returned.Also the threads of the screw were found stripped.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND SCREW LOCK D4.5X36MM
Type of Device
DELTA XTEND IMPLANTS : SCREWS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key16453051
MDR Text Key310351950
Report Number1818910-2023-04672
Device Sequence Number1
Product Code HSD
UDI-Device Identifier10603295027850
UDI-Public10603295027850
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
K192855
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number130790036
Device Catalogue Number130790036
Device Lot Number5422067
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/28/2023
Initial Date FDA Received02/28/2023
Supplement Dates Manufacturer Received02/28/2023
Supplement Dates FDA Received03/01/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/22/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
DXTEND SCREW LOCK D4.5X36MM; DXTEND SCREW LOCK D4.5X36MM
-
-