Model Number 384539 |
Device Problem
Crack (1135)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The sample is indicated as available for return.As of the date of this report, the sample has not been returned.A follow-up report will be provided once the device has been received and reviewed.
|
|
Event Description
|
¿less than 24 hours after insertion, the bedside nurse noted leakage.Upon assessment, the picc had a crack in the plastic part of the hub.We replaced it with another brand, but still have the broken one.¿.
|
|
Event Description
|
¿less than 24 hours after insertion, the bedside nurse noted leakage.Upon assessment, the picc had a crack in the plastic part of the hub.We replaced it with another brand, but still have the broken one.¿.
|
|
Manufacturer Narrative
|
A review of the dhr and inspection records was conducted, and no similar concerns were found.After three notifications, there has been no sample returned for review.Additionally, no photographic or visual evidence of any kind was provided which would have allowed for the allegation to be reviewed.Without such evidence, the results are inconclusive and determining a definite root cause and corrective action is not possible.If the sample is returned at a future date, this complaint will be reopened for further evaluation at that time.
|
|