RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Calibration Problem (2890)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/01/2023 |
Event Type
malfunction
|
Event Description
|
Philips received a complaint by the customer on the v60 indicating that the device has black screen inoperable, code of 1006: data acquisition pcba failure.It was reported there was no patient involvement at the time the issue was discovered.The unit was outside of clinical use; there was no report of harm.The customer evaluated the device with the assistance of the remote service engineer (rse) and confirmed the reported problem in the device history log.The customer replaced the data acquisition (da) pcba to motor controller pcba cable, and da pcba.The device passed required performance verification tests per philips standards and was returned to service.
|
|