|
Catalog Number 05.001.080 |
Device Problems
Break (1069); Unintended System Motion (1430); Mechanical Jam (2983)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi:(b)(4).
|
|
Event Description
|
It was reported from colombia that the pen drive device after activating, it is direct.It was also reported that device continued to work because the valve is faulty and remains blocked.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure, or if a spare device was available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown but was noted to have occurred in 2023.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the device had unintended activation/motion, insufficient/low power, air leak and component damage.It was further found that the device failed pre-test for general condition, check for unintended motion, check handpiece in the "lock" mode, check general function with the apd handswitch, check power with the test bench, check speed with the tachometer, check for internal leak tightness and check for external leak tightness.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to component wear.
|
|
Search Alerts/Recalls
|
|
|