• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-53
Device Problem Circuit Failure (1089)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/21/2023
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during preventive maintenance (pm) performed by a getinge service territory manager (stm) , the cs300 intra-aortic balloon pump (iabp) unit had a solenoid driver board issue.There was no patient involvement.
 
Manufacturer Narrative
Updated fields: b4, d9(device available for eval), g3, g6, h2, h3, h4, h6(investigation type, investigation findings, component codes & investigation conclusions), h10.The fse that encountered the issue evaluated the unit and found solenoid driver board was out of the correct voltage range.There were no logs available as they had been cleared before this issue was discovered.The fse attempted to boost the voltage using board switches.To resolve the issue the fse replaced the solenoid driver board (0670-00-0639).Full pm, safety, calibration, and functionality checks passed factory specifications.Returned to clinical service upon completion.The maquet failure analysis and testing dept.(fat) received part number 0670-00-0639 pcb,solenoid driver with a reported unit failure of solenoid driver board has low voltage readings.The failure analysis and testing dept.Installed part number 0670-00-0639 pcb,solenoid driver into the cs 300 test fixture and tested the board to factory specifications per the cs300 service manual.The fat was able to replicate the failure of low voltage readings when the calibration procedure for the board was performed.The board failed testing.Part no longer able to be tested by the supplier.Investigation is complete.Retaining the part in the failure analysis and testing department per procedure.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key16469275
MDR Text Key310529400
Report Number2249723-2023-01355
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 02/21/2023
Initial Date FDA Received03/02/2023
Supplement Dates Manufacturer Received04/15/2024
Supplement Dates FDA Received04/15/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/02/2005
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
N/A.
-
-